Chapters Transcript Video WavelinQ™ EndoAVF System Product Demonstration Animation depicts WavelinQ™ EndoAVF catheter delivery, alignment and activation for the creation of an endovascular fistula. After venous and arterial access have been achieved and guide wires are in place. Advance the arterial catheter over the guide wire and through the arterial sheath into the brachial artery. Under fluoroscopic guidance, advance the catheter to the target endo A VF site. Next, rotate the arterial catheter until the illumination of the rotational indicators appear as open boxes and the peaks of the backstop are pointed towards the target vein. Under fluoroscopic guidance, advance the venous catheter toward the target endo af creation site. Once correctly aligned, advance the venous catheter until the arc of the electrode sits within the concave surface of the backstop. On the arterial catheter. The electrode should appear compressed under fluoroscopic visualization press and hold the yellow cut button on the electro surgical pencil to create the channel between the artery and vein. After activation, remove the venous catheter, followed by the arterial catheter. Published September 5, 2024 Created by